52551 Magmeter
5. Wiring
5.1 Basic Wiring
2551 Magmeter
4.5-26.4
VDC
Relay 1
Relay 2
NO1
NO2
NO3
NC1
NC2
NC3
C1
C2
C3
Relay 3
V+
V-
+ VDC
- VDC
Aux Pwr
3
Frequency Out
Frequency ÷ 10 Out
(5 to 26.4 VDC)
**********************
Digital (S L) Out
(4.5 to 6.5 VDC)
1
2
3
4
Auxiliary Power for:
Display backlighting
Relay Coils
Ground
Signal
1
2
3
4
5-24 VDC
Ground
Not used
Open Collector Out
2551 Magmeter
4 to 20 mA output
The 3-2551-22 and -42 Magmeters provide a passive 4 to 20 mA
loop output.
External loop power (24 VDC) is required.
Factory standard calibration is 4 - 20 mA = 0 - 5 m/s.
The 4 to 20 mA output can be spanned to any range, from
-10 m/s to +10 m/s.
AUX PWR must be connected to power the display backlighting
and to power to the relay coils if included.
5.2 Wiring: Mirror Relay 1 output
The 3-2551-21 and -41 Magmeter can be con gured in the
OPTIONS menu to provide an Open Collector output in lieu of
the sensor signal provided by Frequency or Digital (S
3
L) output
selections. The Open Collector output can be programmed via the
Relay 1 menu.
Frequency output
3-2551-21 and -41 Magmeters may be programmed in the
OPTIONS menu to provide an open collector FREQUENCY
output.
The maximum frequency output is 1000 Hz (@ 10 meters
per second)
If the Frequency ÷ 10 output is selected, the maximum
frequency is 100 Hz (@ 10 meters per second).
AUX PWR must be connected to power the display backlighting
and to power the relay coils if included.
5.3 2551 and other manufacturer's instruments
When using the 2551 in a system with other manufacturer's
equipment, a 10 K pull-up resistor (not supplied) may be
required to power the open collector output.
1
2
3
4
Frequency Out
2551 Magmeter
24 VDC
NO3
NC3
C3
Relay 3
V+
V-
5-24 VDC
Ground
Ground
Aux Pwr
Sensr Gnd
(SHIELD)
Sensr IN
(RED)
Non-Signet
Flow Instrument
Ground
Frequency Input
5-24 VDC Out
Auxiliary Power for:
Display backlighting
Relay Coils
10 KΩ
Pull-up Resistor
2551 Magmeter
Relay 1
Relay 2
NO1
NO2
NO3
NC1
NC2
NC3
C1
C2
C3
Relay 3
V+
21.6 to
26.4 VDC
V-
+ VDC
- VDC
Aux Pwr
4-20 mA Loop Output
1
2
3
4
Auxiliary Power for:
Display backlighting
Relay Coils
4
20