2610 DO Sensor
1. Wire the 9900 and 3-2610-41 as shown in the diagram.
NOTE: The wiring of the 3-2610-41 is nonstandard:
• 3-2610-41 Red wire is connected to 12 to 24 VDC
• 3-2610-41 White wire is connected to S
3
L Data
• 3-2610-41 Black wire is connected to VDC Ground
• A jumper wire must be connected between the VDC
Ground and S
3
L Gnd.
2. On the 9900 press and hold the ENTER key for 2
seconds.
3. Press the key to select the INPUT menu item.
4. Press the Enter key to access the INPUT menus.
5. Press the key to select the TYPE menu item.
6. Press the key to change the
TYPE selection.
7. Enter the code.
8. Repeatedly press the key until the “4-20 mA INPUT”
menu item is ashing and then press ENTER.
9. The 9900 will prompt “All settings will be reset”. Press the key to select “Yes” and then press ENTER.
10. The 9900 will return to the View Mode display.
11. Press and hold the ENTER key for 2 seconds.
12. Press the key to select the INPUT menu item.
13. In the NAME menu item, press the key to change the displayed name from “4-20 mA INPUT” to a more
descriptive name (e.g., "DISSOLVED O2") and press ENTER when done.
14. Press the key to select SENSOR UNIT menu item.
15. Press the to change the label from UNIT to MG/L or PPM and press ENTER.
16. Press the key and ensure the 4 mA VALUE is set to 0.0000.
17. Press the key and change the 20 mA VALUE from 5.0000 to 20.000 and press ENTER.
18. Press both the and keys simultaneously to return to the Menu.
19. Press the to select the LOOP menu and press ENTER.
20. Set the 4 MA SET POINT to your desired value. The 2610 is factory set for a 0 to 20 mg/L output. Press
ENTER when done.
21. Press the key to select the 20 MA SET POINT and set to your desired value. The 2610 is factory set for a 0
to 20 mg/L output. Press ENTER when done.
22. Press both the and keys simultaneously to return to the Menu.
23. Press the key twice to select the OPTION menu and press ENTER.
24. Press the key twice to select the SET BAR MIN option. Change this option if desired. The 2610 is factory
set for a 0 to 20 mg/L output. Press ENTER when done.
25. Press the key to select the SET BAR MAX option. Change this option if desired. The 2610 is factory set for
a 0 to 20 mg/L output. Press ENTER when done.
26. Press both the and keys simultaneously to return to the Menu.
27. Enter the other menus and set up the unit as desired for your application.
28. Press both the and keys simultaneously to return to the View Menu.
Wiring and Set Up
3-2610-41 to 9900 Set Up
The 3-2610-41 Optical Dissolved Oxygen Sensor with S
3
L was designed to emulate a 4 to 20 mA current input
device, 8058, on the 9900 controller. This allows the 3-2610-41 to be backward compatible with all existing 9900
controllers.
RED
RED
BLACK
BLACK
WHITE
SHIELD
PWR–
DATA
GND
SHLD
V+
9900 S
3
L Inputs
9900 Power Inputs
LOOP–
LOOP+
PWR+
RED
Blue
Green
Brown
3-2610-41
No connection
V+
V-
24 VDC
Regulated
Power Supply
9900 Generation III supports 3-2610-41 direct connection.
Please refer to the 9900 product manual for setup instructions.